This shows you the differences between two versions of the page.
| Both sides previous revision Previous revision Next revision | Previous revision | ||
|
textbook:nrctextbook:chapter8 [2025-04-16 15:00] Merja Herzig |
textbook:nrctextbook:chapter8 [2025-09-01 13:46] (current) Merja Herzig |
||
|---|---|---|---|
| Line 6: | Line 6: | ||
| ### | ### | ||
| There are two principal means to measure radionuclide activities: radiometric and mass | There are two principal means to measure radionuclide activities: radiometric and mass | ||
| - | spectrometric. Radiometric methods are based on detection and measurement of radiation emitted by radionuclides whereas in mass spectrometric methods number of atoms are counted. The results of these methods, [[textbook: | + | spectrometric. Radiometric methods are based on detection and measurement of radiation emitted by radionuclides whereas in [[textbook: |
| ### | ### | ||
| Line 61: | Line 61: | ||
| ;;# | ;;# | ||
| - | {{: | ||
| + | {{: | ||
| Figure VIII.1. Effect of counting geometry on radiation detection of a point source. | Figure VIII.1. Effect of counting geometry on radiation detection of a point source. | ||
| ### | ### | ||
| - | In practice the situation is more complicated since the sources are seldom point sources. As a rule the geometry factor is the higher the closer is the source to the detector. To improve geometry in [[textbook: | + | In practice the situation is more complicated since the sources are seldom point sources. As a rule the geometry factor is the higher the closer is the source to the detector. To improve geometry in [[textbook: |
| ### | ### | ||
| Line 86: | Line 86: | ||
| - | {{: | + | {{: |
| Figure VIII.2. Observed count rate (R) as a function of count rate taking into account 10 µs dead-time of the detector. | Figure VIII.2. Observed count rate (R) as a function of count rate taking into account 10 µs dead-time of the detector. | ||
| Line 138: | Line 138: | ||
| - | {{: | + | {{: |
| Figure VIII.3. Gamma spectrum of < | Figure VIII.3. Gamma spectrum of < | ||
| Line 171: | Line 171: | ||
| * A single channel analyzer (SCA) counts only pulses at a defined height range. As described above, selection of pulse height range is accomplished with voltage discriminators, | * A single channel analyzer (SCA) counts only pulses at a defined height range. As described above, selection of pulse height range is accomplished with voltage discriminators, | ||
| - | {{: | + | {{: |
| Figure VIII.4. Components and scheme of radiation measurement equipment systems. PMT is [[textbook: | Figure VIII.4. Components and scheme of radiation measurement equipment systems. PMT is [[textbook: | ||
| {{anchor: | {{anchor: | ||
| + | {{anchor: | ||
| ===== 8.4. Energy resolution ===== | ===== 8.4. Energy resolution ===== | ||
| Line 184: | Line 185: | ||
| - | {{: | + | {{: |
| Figure VIII.5. Energy resolution of spectrum peak. | Figure VIII.5. Energy resolution of spectrum peak. | ||
| Line 235: | Line 236: | ||
| * [[textbook: | * [[textbook: | ||
| *[[textbook: | *[[textbook: | ||
| + | |||
| + | {{anchor: | ||
| ===== 8.6. Measurement of radionuclides with mass spectrometry ===== | ===== 8.6. Measurement of radionuclides with mass spectrometry ===== | ||
| ### | ### | ||
| - | An alternative for radiometric methods for the determination of radionuclide activities is mass spectrometry. Most often inductively-coupled mass spectrometry (ICP-MS) is today used for this purpose. As mentioned, mass spectrometer counts atoms instead of radiation. This makes mass spectrometry particularly suitable for the measurement of long-lived radionuclides for which the detection limit of mass spectrometry is below those obtained by radiometric methods. For example, | + | An alternative for radiometric methods for the determination of radionuclide |
| - | for < | + | achieved by ICP-MS. In principle all radionuclides with [[textbook: |
| - | achieved by ICP-MS. In principle all radionuclides with half-lives longer than 100 years can be measured by ICP-MS. However, for the radionuclides with half-lives round this limit, radiometric | + | |
| ### | ### | ||
| ### | ### | ||
| - | The components of an ICP-MS are presented in Figure VIII.6. The sample solution is introduced into the system by a nebulizer which turns the solution into a fine mist (aerosol). This is transferred with argon flow into the torch where plasma is created with the help of radiofrequency. Plasma atomizes the sample, ionizes the atoms and the ions are directed into a mass analyzer for the separation of ions based on their mass to charge ratio (//m/z//). | + | The components of an ICP-MS are presented in [[textbook: |
| ### | ### | ||
| + | {{anchor: | ||
| {{: | {{: | ||
| Line 256: | Line 258: | ||
| ### | ### | ||
| - | The mass analyzer is either quadrupole or double focusing system. The former is smaller, cheaper and easier to operate. The latter, however, is much more sensitive yielding to lower detection limits and to more accurate isotopic ratios. A quadrupole consists of four metallic rods aligned in a parallel diamond pattern. By placing a direct current field on one pair of opposite rods and a radio frequency field on the other pair, an ion of a selected mass and charge ratio (//m/z//) is allowed to pass | + | The mass analyzer is either quadrupole or double focusing system. The former is smaller, cheaper and easier to operate. The latter, however, is much more sensitive yielding to lower detection limits and to more accurate |
| through the rods to the detector while the others are forced out of this path. By varying the combinations of voltages and frequency, an array of different //m/z// ratio ions can be scanned in a very short time. The high-resolution double focusing system in turn consists of an electromagnet and an electrostatic analyzer in series. After mass separation the ions are detected and counted. | through the rods to the detector while the others are forced out of this path. By varying the combinations of voltages and frequency, an array of different //m/z// ratio ions can be scanned in a very short time. The high-resolution double focusing system in turn consists of an electromagnet and an electrostatic analyzer in series. After mass separation the ions are detected and counted. | ||
| ### | ### | ||
| Line 262: | Line 264: | ||
| ### | ### | ||
| - | Some radionuclides, | + | Some [[textbook: |
| - | compared to < | + | |
| nevertheless much higher than that of plutonium due to the greater abundances of the polyatomic ion forming elements, in this case Hg, Cl and U. Therefore, chemical separations are needed to enable measurement of radionuclides at very low concentrations. Third type of interference comes from broadening the neighbor mass peaks at higher concentration. Due to this, for example, < | nevertheless much higher than that of plutonium due to the greater abundances of the polyatomic ion forming elements, in this case Hg, Cl and U. Therefore, chemical separations are needed to enable measurement of radionuclides at very low concentrations. Third type of interference comes from broadening the neighbor mass peaks at higher concentration. Due to this, for example, < | ||
| much higher concentrations compared to plutonium causes extra counts to the mass peak of < | much higher concentrations compared to plutonium causes extra counts to the mass peak of < | ||
| Line 270: | Line 271: | ||
| ### | ### | ||
| - | ICP-MS is increasingly used for the measurement of long-lived radionuclides, | + | ICP-MS is increasingly used for the measurement of [[textbook: |
| - | measured by mass spectrometry due to interference of uranium. Thus, to determine all relevant plutonium isotopes both mass and alpha spectrometry are needed. | + | |
| ### | ### | ||
email: mst@evalion.cz | tel: +420 224 358 331 | Copyright © 2021 A-CINCH
This project has received funding from the Euratom research and training programme 2019–2020 under grant agreement No. 945301.