This shows you the differences between two versions of the page.
| Both sides previous revision Previous revision Next revision | Previous revision | ||
|
textbook:nrctextbook:chapter8 [2025-04-16 13:57] Merja Herzig |
textbook:nrctextbook:chapter8 [2025-09-01 13:46] (current) Merja Herzig |
||
|---|---|---|---|
| Line 6: | Line 6: | ||
| ### | ### | ||
| There are two principal means to measure radionuclide activities: radiometric and mass | There are two principal means to measure radionuclide activities: radiometric and mass | ||
| - | spectrometric. Radiometric methods are based on detection and measurement of radiation emitted by radionuclides whereas in mass spectrometric methods number of atoms are counted. The results of these methods, [[textbook: | + | spectrometric. Radiometric methods are based on detection and measurement of radiation emitted by radionuclides whereas in [[textbook: |
| ### | ### | ||
| Line 61: | Line 61: | ||
| ;;# | ;;# | ||
| - | {{: | ||
| + | {{: | ||
| Figure VIII.1. Effect of counting geometry on radiation detection of a point source. | Figure VIII.1. Effect of counting geometry on radiation detection of a point source. | ||
| ### | ### | ||
| - | In practice the situation is more complicated since the sources are seldom point sources. As a rule the geometry factor is the higher the closer is the source to the detector. To improve geometry in [[textbook: | + | In practice the situation is more complicated since the sources are seldom point sources. As a rule the geometry factor is the higher the closer is the source to the detector. To improve geometry in [[textbook: |
| ### | ### | ||
| Line 86: | Line 86: | ||
| - | {{: | + | {{: |
| Figure VIII.2. Observed count rate (R) as a function of count rate taking into account 10 µs dead-time of the detector. | Figure VIII.2. Observed count rate (R) as a function of count rate taking into account 10 µs dead-time of the detector. | ||
| Line 138: | Line 138: | ||
| - | {{: | + | {{: |
| Figure VIII.3. Gamma spectrum of < | Figure VIII.3. Gamma spectrum of < | ||
| Line 148: | Line 148: | ||
| ### | ### | ||
| - | If the sample contains several [[textbook: | + | If the sample contains several [[textbook: |
| - | which positions of peaks and their relative intensities are made use of. Alpha spectrometry with semiconductor detectors is used to measure both activities of alpha emitters and their isotopic ratios, the latter bringing often valuable information, | + | |
| - | however, only one beta emitter is measured at a time. Sometimes it is possible to measure two beta emitters from the same samples if the energy difference of the two beta emissions is high enough. | + | |
| ### | ### | ||
| Line 159: | Line 157: | ||
| Radiation measurement equipment systems consist of the following components (Figure VIII.4): | Radiation measurement equipment systems consist of the following components (Figure VIII.4): | ||
| - | * Detector, the function of which is to transform the energy of radiation into an electric pulse (gas ionization detectors and semiconductor detectors) or to a light pulse (scintillation detectors). Various detectors are discussed in later chapters in more detail. | + | * Detector, the function of which is to transform the energy of radiation into an electric pulse ([[textbook: |
| * Voltage source which collects the initial electric pulses into electrodes. | * Voltage source which collects the initial electric pulses into electrodes. | ||
| * Preamplifier which amplifies the weak pulses coming from semiconductor detectors to enable the pulse transfer through cables into the main amplifier. | * Preamplifier which amplifies the weak pulses coming from semiconductor detectors to enable the pulse transfer through cables into the main amplifier. | ||
| * Main amplifier is called linear amplifier since its function is not only to increase the pulse height to a measureable one but also preserve the energy information. This is done by amplifying each initial pulse with the same factor so that the observed pulse heights are linearly related to the heights of the initial pulses coming from the detector and the preamplifier. | * Main amplifier is called linear amplifier since its function is not only to increase the pulse height to a measureable one but also preserve the energy information. This is done by amplifying each initial pulse with the same factor so that the observed pulse heights are linearly related to the heights of the initial pulses coming from the detector and the preamplifier. | ||
| - | * In scintillation detectors there is, instead of preamplifier and linear amplifier, a photomultiplier tube (PMT) which converts the light pulse into an electric pulse and amplifies the initial pulse into a measureable | + | * In scintillation detectors there is, instead of preamplifier and linear amplifier, a [[textbook: |
| ### | ### | ||
| - | After this there are two options depending on whether the equipment is used as a multichannel analyzer or as a single-channel analyzer. The former is used in energy spectrometry and the latter in pulse counting. | + | After this there are two options depending on whether the equipment is used as a [[textbook: |
| ### | ### | ||
| {{anchor: | {{anchor: | ||
| - | * A multichannel analyzer (MCA) sorts the pulses into various channels depending on their pulse height which is proportional to the energy of the particle or ray. For example 1 mV pulse goes to channel 1, 12 mV pulse to channel 12 and 715 mV pulse to channel 715. This results in the formation of an energy spectrum. A multichannel analyzer may have even thousands of channels. Prior to sorting the pulses into channels the analog-to-digital converter (ADC) transforms the analogical pulses into digital form. | + | * A multichannel analyzer (MCA) sorts the pulses into various channels depending on their pulse height which is proportional to the energy of the particle or ray. For example 1 mV pulse goes to channel 1, 12 mV pulse to channel 12 and 715 mV pulse to channel 715. This results in the formation of an energy spectrum. A multichannel analyzer may have even thousands of channels. Prior to sorting the pulses into channels the analog-to-digital converter (ADC) transforms the analogical pulses into digital form. {{anchor: |
| - | * A single channel analyzer (SCA) counts only pulses at a defined height range. As described above, selection of pulse height range is accomplished with voltage discriminators, | + | * A single channel analyzer (SCA) counts only pulses at a defined height range. As described above, selection of pulse height range is accomplished with voltage discriminators, |
| - | {{: | + | {{: |
| - | Figure VIII.4. Components and scheme of radiation measurement equipment systems. PMT is photomultiplier tube. | + | Figure VIII.4. Components and scheme of radiation measurement equipment systems. PMT is [[textbook: |
| {{anchor: | {{anchor: | ||
| + | {{anchor: | ||
| ===== 8.4. Energy resolution ===== | ===== 8.4. Energy resolution ===== | ||
| ### | ### | ||
| In energy spectrometry it is essential that the measurement system can differentiate different particle or ray energies as efficiently as possible. This is mostly dependent on the type of detector. The better the energy resolution the better the system can differentiate energies close to each other and the narrower are the observed peaks in a spectrum. Resolution (//R//) is expressed as the peak width at half of the height of peak maximum (FWHM = full width at half maximum) (Figure | In energy spectrometry it is essential that the measurement system can differentiate different particle or ray energies as efficiently as possible. This is mostly dependent on the type of detector. The better the energy resolution the better the system can differentiate energies close to each other and the narrower are the observed peaks in a spectrum. Resolution (//R//) is expressed as the peak width at half of the height of peak maximum (FWHM = full width at half maximum) (Figure | ||
| - | VIII.5). Instead of the absolute value the energy resolution can also expressed as the relative value // | + | VIII.5). Instead of the absolute value the energy resolution can also expressed as the relative value // |
| - | as the FWHM of the < | + | |
| - | 100 keV (5%) for 2 MeV rays, 70 keV (7%) for 1 MeV rays and about 50 keV (10%) for 0.5 MeV rays. Thus the germanium detectors have about 50-times better resolution compared to the NaI(Tl) detectors. Silicon semiconductor alpha detectors have resolutions between 20-30 keV (0.4-0.6% for typical alpha energies of 4-7 MeV) which are about ten times lower than values obtainable with liquid scintillation counters. | + | |
| ### | ### | ||
| - | {{: | + | {{: |
| Figure VIII.5. Energy resolution of spectrum peak. | Figure VIII.5. Energy resolution of spectrum peak. | ||
| Line 201: | Line 198: | ||
| 1. Gas ionization detectors | 1. Gas ionization detectors | ||
| - | * Ionisation | + | * [[textbook: |
| - | * Proportional counter | + | * [[textbook: |
| - | * Geiger-Muller (GM) tube | + | * [[textbook: |
| 2. Scintillation detectors | 2. Scintillation detectors | ||
| - | * Liquid scintillation | + | * [[textbook: |
| - | * Solid scintillation detectors | + | * [[textbook: |
| - | 3. Semiconductor detectors | + | 3. Semiconductor detectors |
| ### | ### | ||
| - | **Alpha radiation** is most accurately measured with semiconductor detectors. Their background pulses are very low and their energy resolution is good, at 5 MeV energies even 15 keV. The energy resolution of liquid scintillation counting, another choice to measure alpha radiation, is about ten times poorer than that of semiconductor detectors. Sample preparation for liquid scintillation counting is, however, clearly less difficult since the sample is just dissolved in the liquid scintillation cocktail for measurement while for semiconductor detectors counting sources need to be prepared by electrodeposition or microcoprecipitation. Another benefit of liquid scintillation | + | **[[textbook: |
| - | counting is a very good, practically 100%, counting efficiency. In liquid scintillation counting the beta emitters present can cause problems by creating extra pulses to alpha peaks. In modern liquid scintillation counters this is overcome by alpha-beta discrimination system that differentiates alpha and beta pulses from each other and count them separately. Due to the poor energy resolution liquid | + | counting is a very good, practically 100%, [[textbook: |
| scintillation counting is not a proper method to determine the isotopic ratios of alpha emitters. For this purpose semiconductor detectors need to be used. | scintillation counting is not a proper method to determine the isotopic ratios of alpha emitters. For this purpose semiconductor detectors need to be used. | ||
| ### | ### | ||
| Line 219: | Line 216: | ||
| ### | ### | ||
| - | For **beta radiation** the most often used method is liquid scintillation counting. It yields into high counting efficiencies and is suitable also for low energy beta radiation. Liquid scintillation counting also enables determination of beta spectra. Usually, however, due to the continuous nature of beta spectra, only one beta emitter can be measured at a time. Other options to measure beta radiation are the gas ionization detectors, GM tube and proportional counters, the latter of which can also | + | For **[[textbook: |
| - | produce beta spectra. The drawbacks of gas ionization detectors are more laborious counting source preparation, | + | |
| ### | ### | ||
| ### | ### | ||
| - | For the measurement of **gamma radiation** solid scintillation and semiconductor detectors are used. The benefit of solid scintillation detectors is their higher counting efficiency as the detectors can be produced in large sizes and they are often of well-type in which the sample is inside the detector.Solid scintillation detectors are usually utilized for gamma counting in single-channel mode. The benefit of semiconductor detectors is their superior energy resolution compared to solid scintillators and therefore they are typically used for gamma spectrometric measurements in radionuclide identifications and measurement of radionuclide activities from samples having several gamma-emitting nuclides. | + | For the measurement of **[[textbook: |
| ### | ### | ||
| Line 234: | Line 230: | ||
| - | * semiconductor detector(s) for gamma spectrometry for the determination of radionuclides from environmental and radioactive waste samples, for example | + | * [[textbook: |
| - | * gamma counter(s) having a solid scintillation detector and a sample changer for the measurement of tracer gamma emitters used in model experiments | + | * gamma counter(s) having a [[textbook: |
| - | * liquid scintillation counter(s) with alpha-beta discrimination for the measurement of tracer beta and alpha emitters as well as beta and alpha emitters separated from various samples | + | * [[textbook: |
| * low background liquid scintillation counter(s) for the measurement of low beta activities | * low background liquid scintillation counter(s) for the measurement of low beta activities | ||
| - | * semiconductor alpha detector(s) for the measurement of alpha emitters separated from various samples | + | * [[textbook: |
| - | * gas ionization detector(s) to measure low activity beta emitters separated from various samples | + | *[[textbook: |
| + | |||
| + | {{anchor: | ||
| ===== 8.6. Measurement of radionuclides with mass spectrometry ===== | ===== 8.6. Measurement of radionuclides with mass spectrometry ===== | ||
| ### | ### | ||
| - | An alternative for radiometric methods for the determination of radionuclide activities is mass spectrometry. Most often inductively-coupled mass spectrometry (ICP-MS) is today used for this purpose. As mentioned, mass spectrometer counts atoms instead of radiation. This makes mass spectrometry particularly suitable for the measurement of long-lived radionuclides for which the detection limit of mass spectrometry is below those obtained by radiometric methods. For example, | + | An alternative for radiometric methods for the determination of radionuclide |
| - | for < | + | achieved by ICP-MS. In principle all radionuclides with [[textbook: |
| - | achieved by ICP-MS. In principle all radionuclides with half-lives longer than 100 years can be measured by ICP-MS. However, for the radionuclides with half-lives round this limit, radiometric | + | |
| ### | ### | ||
| ### | ### | ||
| - | The components of an ICP-MS are presented in Figure VIII.6. The sample solution is introduced into the system by a nebulizer which turns the solution into a fine mist (aerosol). This is transferred with argon flow into the torch where plasma is created with the help of radiofrequency. Plasma atomizes the sample, ionizes the atoms and the ions are directed into a mass analyzer for the separation of ions based on their mass to charge ratio (//m/z//). | + | The components of an ICP-MS are presented in [[textbook: |
| ### | ### | ||
| + | {{anchor: | ||
| {{: | {{: | ||
| Line 261: | Line 258: | ||
| ### | ### | ||
| - | The mass analyzer is either quadrupole or double focusing system. The former is smaller, cheaper and easier to operate. The latter, however, is much more sensitive yielding to lower detection limits and to more accurate isotopic ratios. A quadrupole consists of four metallic rods aligned in a parallel diamond pattern. By placing a direct current field on one pair of opposite rods and a radio frequency field on the other pair, an ion of a selected mass and charge ratio (//m/z//) is allowed to pass | + | The mass analyzer is either quadrupole or double focusing system. The former is smaller, cheaper and easier to operate. The latter, however, is much more sensitive yielding to lower detection limits and to more accurate |
| through the rods to the detector while the others are forced out of this path. By varying the combinations of voltages and frequency, an array of different //m/z// ratio ions can be scanned in a very short time. The high-resolution double focusing system in turn consists of an electromagnet and an electrostatic analyzer in series. After mass separation the ions are detected and counted. | through the rods to the detector while the others are forced out of this path. By varying the combinations of voltages and frequency, an array of different //m/z// ratio ions can be scanned in a very short time. The high-resolution double focusing system in turn consists of an electromagnet and an electrostatic analyzer in series. After mass separation the ions are detected and counted. | ||
| ### | ### | ||
| Line 267: | Line 264: | ||
| ### | ### | ||
| - | Some radionuclides, | + | Some [[textbook: |
| - | compared to < | + | |
| nevertheless much higher than that of plutonium due to the greater abundances of the polyatomic ion forming elements, in this case Hg, Cl and U. Therefore, chemical separations are needed to enable measurement of radionuclides at very low concentrations. Third type of interference comes from broadening the neighbor mass peaks at higher concentration. Due to this, for example, < | nevertheless much higher than that of plutonium due to the greater abundances of the polyatomic ion forming elements, in this case Hg, Cl and U. Therefore, chemical separations are needed to enable measurement of radionuclides at very low concentrations. Third type of interference comes from broadening the neighbor mass peaks at higher concentration. Due to this, for example, < | ||
| much higher concentrations compared to plutonium causes extra counts to the mass peak of < | much higher concentrations compared to plutonium causes extra counts to the mass peak of < | ||
| Line 275: | Line 271: | ||
| ### | ### | ||
| - | ICP-MS is increasingly used for the measurement of long-lived radionuclides, | + | ICP-MS is increasingly used for the measurement of [[textbook: |
| - | measured by mass spectrometry due to interference of uranium. Thus, to determine all relevant plutonium isotopes both mass and alpha spectrometry are needed. | + | |
| ### | ### | ||
email: mst@evalion.cz | tel: +420 224 358 331 | Copyright © 2021 A-CINCH
This project has received funding from the Euratom research and training programme 2019–2020 under grant agreement No. 945301.